Specifications
Light Source: UV300W Bulb
Lamp Life: 1,000 Hours
Temperature Range: Room Temperature~80 °C
Timer Range: 0~9,999 H
Turntable Diameter: φ300 Mm
Turntable Speed: 3±1 Rpm
Heating Mode: Forced Hot-air Circulation
Insulation Layer: Glass Fiber
Drive Motor: 1/4 HP
Inner Dimensions: 50 × 50 × 60 Cm
Overall Dimensions: 100 × 65 × 117 Cm
Net Weight: 135 Kg
Highlight:

ASTM D1148 UV Anti-Yellowing Aging Test Chamber

,

UV300W Lamp UV Aging Test Chamber

,

Forced Hot-Air Circulation Yellowing Resistance Tester

Basic Infomation
Place of Origin: CHINA
Brand Name: YOUBI
Certification: CE,ISO
Model Number: UP-6199
Payment & Shipping Terms
Packaging Details: Professional Export Wooden Box Packaging
Delivery Time: 5-15 work days
Payment Terms: T/T,D/P,Western Union,Paypal
Supply Ability: 30 Unit/month
Product Description
Product Overview

The UV Anti-Yellowing Aging Test Chamber — also known as a bulb-type UV discoloration tester or yellowing resistance tester — determines how well white and light-colored materials resist yellowing under simulated sunlight and heat.

A 300 W UV bulb irradiates specimens placed on a slowly rotating φ300 mm turntable, while a forced hot-air circulation system holds the chamber at a set temperature up to 80 °C. After the programmed exposure period, the exposed specimen is compared with an unexposed control (or a standard grey scale) to grade the degree of discoloration.

Typical test materials include shoe soles, leather, PU, rubber, fabrics and other light-colored polymer or coated substrates.

UP-6199 UV Anti-Yellowing Aging Test Chamber for shoe material, plastic and rubber yellowing resistance testing, compliant with ASTM D1148 and HG/T3689
Main Features
Uniform, Repeatable Exposure

The φ300 mm perforated stainless-steel sample disc rotates at 3 ± 1 rpm so that every specimen receives the same UV dose, removing the "hot spot" bias of static sample placement.

Perforated φ300 mm stainless steel rotating sample turntable at 3±1 rpm inside the UV anti-yellowing test chamber
Stable Thermal Environment

A top-mounted 300 W UV lamp with reflector works together with forced hot-air circulation and perforated SUS304 inner walls, giving even temperature distribution across the 50 × 50 × 60 cm working volume.

SUS304 stainless steel inner chamber with top-mounted 300 W UV lamp, reflector, wire shelf and hot-air circulation perforations
Built for Continuous Laboratory Use

Mirror-polished SUS304 stainless-steel interior resists corrosion and wipes clean easily; glass-fiber insulation limits heat loss; the baked-enamel steel exterior keeps the cabinet looking new. The unit is fitted with a front observation window, a side control cluster (temperature controller, LED timer, indicators, ammeter and main switch) and heavy-duty caster wheels for easy repositioning.

Safety Built In

An EGO over-temperature protection device cuts power to the heating circuit and triggers an indicator if the set limit is exceeded, while a safety overload switch with ammeter protects the electrical circuit during long unattended runs.

Flexible Test Setup

Light intensity is adjustable, and the 0–9,999 h LED timer supports both short QC screening cycles and extended aging programs.

Technical Specifications
Item Specification
Product typeBulb-type UV anti-yellowing / discoloration aging test chamber
Light sourceUV300W bulb (sunlight-simulation lamp)
Lamp service life1,000 hours
Test material discφ30 cm (φ300 mm) perforated rotating turntable, 3 ± 1 rpm
Light intensityAdjustable
Temperature rangeRoom temperature ~ 80 °C
Heating modeForced hot-air circulation
TimerLED display, 0 ~ 9,999 h
Insulation layerGlass fiber
Chamber materialInterior: SUS304 stainless steel plate; exterior: baked-enamel / painted steel
Safety protectionEGO over-temperature power-off indicator; safety overload switch with ammeter
Drive motor1/4 HP
Inner box dimensions50 × 50 × 60 cm
Overall dimensions100 × 65 × 117 cm
Net weight135 kg
Power requirementsSingle-phase (1Ø), AC 220 V, 15 A
Test Standards

ASTM D1148, ASTM D2438, JIS-K6301

Note: These methods are commonly referenced for evaluating UV/heat-induced discoloration of white or light-colored surfaces. Please state the exact standard and revision required by your laboratory when requesting a quotation so the configuration can be matched accordingly.

Operation & Maintenance Notes
  1. Replace the UV bulb after 1,000 hours of accumulated operation to keep irradiance consistent.
  2. Do not look directly at the lit UV lamp; keep the door closed during a running cycle.
  3. Keep the perforated chamber walls and turntable clear of debris so hot air can circulate freely.
  4. Verify a single-phase AC 220 V / 15 A supply before installation.
  5. Grade and record specimens promptly after exposure, using a consistent reference sample and lighting conditions for comparable results.
UV Anti-Yellowing Aging Test Chamber for shoe material, plastic and rubber yellowing resistance testing, compliant with ASTM D1148 and HG/T3689
Control panel of the UV Anti-Yellowing Aging Test Chamber with temperature controller, timer and ammeter for anti-yellow aging testing
Frequently Asked Questions
Q: Which materials can be tested?

A: Any white or light-colored material where yellowing is a concern — shoe soles, leather, PU/PVC, rubber, fabrics, coated plastics, adhesives and light-colored finished parts.

Q: Is this a full UV weathering chamber?

A: No. This is a bulb-type chamber configured for rapid UV + heat discoloration (yellowing) screening. It is not equipped with humidity, condensation or water-spray functions; for gloss loss, cracking or moisture-weathering studies, a full weathering chamber is required.

Q: How often should the lamp be replaced?

A: The UV300W bulb has a rated working life of 1,000 hours. We recommend logging operating hours and replacing it at that interval.

Q: Can the light intensity be adjusted?

A: Yes. Light intensity is adjustable so exposure conditions can be matched to different internal or standard test protocols. Specify your required setting when enquiring.

Q: What power supply is needed?

A: Single-phase AC 220 V, 15 A. Please confirm your local supply when placing the order.

Q: Can I order spare bulbs or a custom configuration?

A: Yes — contact us with your test standard, sample size, required irradiance and power supply, and we will confirm the configuration, spare UV bulb kit, packaging and delivery schedule.

Recommended Products
Get in touch with us
Contact Person : Jessica
Tel : 008613422863552
Characters Remaining(20/3000)