The UP-6118 thermal shock test chamber evaluates the endurance of material structures and composite materials under continuous, instantaneous transitions between extremely high and low temperatures. It exposes chemical changes or physical damage caused by thermal expansion within the shortest possible time, making it suitable for universities, research institutions, and quality testing across electronic appliances, LED lighting fixtures, automotive accessories, computer peripherals, plastics and rubber, and similar materials. Test results serve as a basis or reference for product improvement.
Built around a compact design concept, the chamber combines pressurizing, vacuuming, and recharging functions in a single machine to save space and simplify operation, repair, and maintenance.

Reliability and screening tests for:
| Item | Specification |
|---|---|
| Model | UP-6118 |
| Applicable standard | IEC 60068-2-14 Test Na (change of temperature with specified transfer time) |
| Structure | 2-zone moving basket / 3-zone stationary specimen with air dampers |
| Specimen exposure range | −65℃ ~ +150℃ (standard); optional −40℃ / −55℃ low, +125℃ / +200℃ high |
| High-temperature reservoir | +80℃ ~ +200℃ |
| Low-temperature reservoir | −75℃ ~ −10℃ |
| Zone transfer time | ≤10 s (2-zone basket travel); ≤30 s (3-zone damper changeover) — switching action only, excluding recovery |
| Temperature recovery | ≤3 min typical, to within ±2 K of setpoint (80–252 L, standard load; 480 L and heavy loads confirmed at FAT) |
| Temperature fluctuation | ≤±0.5℃ (stabilized reservoirs) |
| Temperature uniformity | ≤±2.0℃ (stabilized reservoirs, 9-point survey) |
| Temperature change method | Instant transfer between pre-conditioned reservoirs; not a linear ℃/min ramp |
| Humidity control | Not applicable — dry-air / N₂ anti-condensation purge optional |
| Control system | Touchscreen PLC, PID + SSR, PT100, multi-cycle, dwell, auto-run, data logging |
| Inner / outer material | SUS304 stainless steel / powder-coated steel |
| Refrigeration | Cascade system for −65℃ low zone, CFC-free refrigerant; air-cooled standard, water-cooled optional |
| Safety devices | Over-temp, compressor, fan, overload, phase loss, power-off, interlock, E-stop |
| Power supply | 3-phase AC 380V ±10%, 50/60 Hz, 3 Φ~ |
| Test volume | 80 L / 150 L / 252 L / 480 L or customized |
| Weight | 600–1500 kg (by model) |
| Warranty | 12 months after shipment, extended by 2 months for ocean transit (14 months total) |


The chamber ships fully assembled. Connect the 3-phase power supply (and cooling water if the water-cooled option is selected), level the cabinet, and run the auto-tune routine. Allow airflow clearance around the condenser and confirm floor loading.
An English operation manual, wiring diagram, and commissioning video are supplied by email.
Yes — we are a professional manufacturer of environmental test chambers, footwear / leather testers, and rubber / plastics testing equipment in Dongguan, China. Every machine carries a 12-month warranty after shipment, extended by 2 months to cover sea transportation. For any issue, contact us by email or phone; we troubleshoot by call or video and provide a confirmed solution within 24–48 hours.
Stock standard models: 3–7 working days. Built-to-order or customized units: 15–20 working days after payment. Urgent schedules can be arranged separately.
Yes. Standard and customized chambers are both available, including OEM/ODM, logo, chamber volume, cable ports, dry-air purge, water cooling, and controller language.
English manuals and videos are provided. We conduct multiple testing runs before shipment and only arrange shipment after the machine is running stably and the data meets the standards.
No. The UP-6118 is a dry-air thermal shock chamber; humidity is not a controlled parameter in Test Na. An optional dry-air / nitrogen purge is offered to prevent condensation on cold-to-hot transitions.
It is the pure transfer / damper-switching time only, from leaving one zone to being fully in the other; temperature recovery is a separate figure.