Thermal Shock Test Chamber UP-6118 with ≤10 s Zone Transfer and −65~+150℃ Range in 2-Zone or 3-Zone Structure

1 Unit
MOQ
Negotiable
Price
Thermal Shock Test Chamber UP-6118 with ≤10 s Zone Transfer and −65~+150℃ Range in 2-Zone or 3-Zone Structure
Features Gallery Product Description Chat Now
Features
Specifications
Humidity Range: 20%~98%RH
Cooling Rate: 1~1.5ºC/min
Heating Rate: 3~5ºC/min
Custom Service: OEM,ODM
Product Name: Thermal Shock Test Chamber
Model NO.: UP-6118
Temperature Range Of High-temperature Tank: 80ºC~200ºC
Temperature Range Of Low-temperature Tank: -10ºC~75ºC
Customized Support: ODM,OEM
Warranty: 1 Year
Exposure Range: −65℃ ~ +150℃
Zone Transfer Time: ≤10 S (2-zone) / ≤30 S (3-zone)
Temperature Recovery: ≤3 Min Typical
Temperature Fluctuation: ≤±0.5℃
Temperature Uniformity: ≤±2.0℃
High-Temp Reservoir: +80℃ ~ +200℃
Low-Temp Reservoir: −75℃ ~ −10℃
Test Volume: 80 L / 150 L / 252 L / 480 L
Power Supply: 3-phase AC 380V ±10%, 50/60 Hz
Inner Material: SUS304 Stainless Steel
Outer Material: Powder-coated Steel
Weight: 600-1500 Kg (by Model)
Specimen Exposure Range: −65℃ ~ +150℃ (standard); Optional −40℃ / −55℃ Low, +125℃ / +200℃ High
High-temperature Reservoir: +80℃ ~ +200℃
Low-temperature Reservoir: −75℃ ~ −10℃
Zone Transfer Time: ≤10 S (2-zone Basket Travel); ≤30 S (3-zone Damper Changeover)
Temperature Recovery: ≤3 Min Typical, To Within ±2 K Of Setpoint
Temperature Fluctuation: ≤±0.5℃ (stabilized Reservoirs)
Temperature Uniformity: ≤±2.0℃ (stabilized Reservoirs, 9-point Survey)
Test Volume: 80 L / 150 L / 252 L / 480 L Or Customized
Power Supply: 3-phase AC 380V ±10%, 50/60 Hz, 3Φ
Highlight:

IEC 60068 thermal shock test chamber

,

Ultra-fast thermal shock chamber ≤10s

,

Wide temperature range test chamber -65℃~+150℃

Basic Infomation
Place of Origin: CHINA
Brand Name: YOUBI
Certification: CE,ISO
Model Number: UP-6118
Payment & Shipping Terms
Packaging Details: Professional Export Wooden Box Packaging
Delivery Time: 5-15 work days
Payment Terms: T/T,D/P,Western Union,Paypal
Supply Ability: 30 Unit/month
Product Description
Product Overview

The UP-6118 thermal shock test chamber evaluates the endurance of material structures and composite materials under continuous, instantaneous transitions between extremely high and low temperatures. It exposes chemical changes or physical damage caused by thermal expansion within the shortest possible time, making it suitable for universities, research institutions, and quality testing across electronic appliances, LED lighting fixtures, automotive accessories, computer peripherals, plastics and rubber, and similar materials. Test results serve as a basis or reference for product improvement.
Built around a compact design concept, the chamber combines pressurizing, vacuuming, and recharging functions in a single machine to save space and simplify operation, repair, and maintenance.

Two-zone thermal shock test chamber, white thermal shock chamber for electronic product temperature cycling test
Applications

Reliability and screening tests for:

  • Semiconductor packages and PCB/PCBA assemblies
  • Automotive electronics and sensors
  • Aerospace components
  • Battery cells and modules
  • Connectors and harnesses
  • Metal parts, plastic and rubber products, and building materials
  • Communication equipment
  • University and third-party laboratory testing
Specifications
ItemSpecification
ModelUP-6118
Applicable standardIEC 60068-2-14 Test Na (change of temperature with specified transfer time)
Structure2-zone moving basket / 3-zone stationary specimen with air dampers
Specimen exposure range−65℃ ~ +150℃ (standard); optional −40℃ / −55℃ low, +125℃ / +200℃ high
High-temperature reservoir+80℃ ~ +200℃
Low-temperature reservoir−75℃ ~ −10℃
Zone transfer time≤10 s (2-zone basket travel); ≤30 s (3-zone damper changeover) — switching action only, excluding recovery
Temperature recovery≤3 min typical, to within ±2 K of setpoint (80–252 L, standard load; 480 L and heavy loads confirmed at FAT)
Temperature fluctuation≤±0.5℃ (stabilized reservoirs)
Temperature uniformity≤±2.0℃ (stabilized reservoirs, 9-point survey)
Temperature change methodInstant transfer between pre-conditioned reservoirs; not a linear ℃/min ramp
Humidity controlNot applicable — dry-air / N₂ anti-condensation purge optional
Control systemTouchscreen PLC, PID + SSR, PT100, multi-cycle, dwell, auto-run, data logging
Inner / outer materialSUS304 stainless steel / powder-coated steel
RefrigerationCascade system for −65℃ low zone, CFC-free refrigerant; air-cooled standard, water-cooled optional
Safety devicesOver-temp, compressor, fan, overload, phase loss, power-off, interlock, E-stop
Power supply3-phase AC 380V ±10%, 50/60 Hz, 3 Φ~
Test volume80 L / 150 L / 252 L / 480 L or customized
Weight600–1500 kg (by model)
Warranty12 months after shipment, extended by 2 months for ocean transit (14 months total)
Thermal shock chamber control panel, touch screen, emergency stop button, industrial environmental test equipment
Inner chamber of thermal shock chamber, open door, stainless steel test room, thermal shock testing equipment
Installation & After-Sales Service

The chamber ships fully assembled. Connect the 3-phase power supply (and cooling water if the water-cooled option is selected), level the cabinet, and run the auto-tune routine. Allow airflow clearance around the condenser and confirm floor loading.
An English operation manual, wiring diagram, and commissioning video are supplied by email.

Frequently Asked Questions
1. Are you a manufacturer? What about warranty and after-sales service?

Yes — we are a professional manufacturer of environmental test chambers, footwear / leather testers, and rubber / plastics testing equipment in Dongguan, China. Every machine carries a 12-month warranty after shipment, extended by 2 months to cover sea transportation. For any issue, contact us by email or phone; we troubleshoot by call or video and provide a confirmed solution within 24–48 hours.

2. What is the delivery term?

Stock standard models: 3–7 working days. Built-to-order or customized units: 15–20 working days after payment. Urgent schedules can be arranged separately.

3. Do you accept customization? Can I add our logo?

Yes. Standard and customized chambers are both available, including OEM/ODM, logo, chamber volume, cable ports, dry-air purge, water cooling, and controller language.

4. How is the machine installed and operated?

English manuals and videos are provided. We conduct multiple testing runs before shipment and only arrange shipment after the machine is running stably and the data meets the standards.

5. Does this chamber control humidity?

No. The UP-6118 is a dry-air thermal shock chamber; humidity is not a controlled parameter in Test Na. An optional dry-air / nitrogen purge is offered to prevent condensation on cold-to-hot transitions.

6. How is "≤10 s" measured, and is it with a load?

It is the pure transfer / damper-switching time only, from leaving one zone to being fully in the other; temperature recovery is a separate figure.

Recommended Products
Get in touch with us
Contact Person : Jessica
Tel : 008613422863552
Characters Remaining(20/3000)