The UP-6195 is a compact benchtop chamber designed for temperature and humidity testing of small components and materials. Its CNC-formed construction gives the unit a clean, modern profile, while the non-reactive handle supports smooth, safe operation. A microcomputer-based controller maintains stable, accurate conditions for repeatable test results.

The chamber is designed to support the following damp-heat and temperature-cycling test methods. Test methods and severity levels are programmed on the controller; conformity to a specific standard depends on the configured test profile.
| Item | Specification |
|---|---|
| Model | UP-6195 |
| Type | Benchtop constant temperature & humidity test chamber |
| Internal volume | 60 L (customizable) |
| Temperature range | -10°C to +150°C (customizable on request) |
| Humidity range | 30% to 95% RH (customizable on request) |
| Temperature fluctuation | ≤ ±0.5°C |
| Temperature uniformity | ≤ ±2°C |
| Control accuracy | ±0.5°C; ±3% RH |
| Heating rate | 1.0–3.0°C/min |
| Cooling rate | 0.7–1.0°C/min |
| Inner chamber material | SUS 304 stainless steel, mirror finish, argon-arc welded |
| Outer chamber material | A3 steel plate, powder-coated |
| Controller | Microprocessor-based temperature & humidity controller |
| Observation window | Double-glazed window with internal lighting |
| Cable / test port | Ø50 mm, located on the left side |
| Safety protection | Independent over-temperature limit alarm with automatic power cut-off |
| Mobility | Lockable PU casters |
Commonly used for environmental simulation and reliability testing of electronic components, PCB/PCBA assemblies, semiconductor devices, automotive sensors and connectors, packaging materials and small material specimens — in electronics manufacturing, automotive supply, research institutes and quality-inspection laboratories.
Chamber volume (60 L standard / customizable), temperature range, humidity range and test-port configuration can be tailored to your required test standard and sample size. When requesting a quotation, please share your test profile, sample dimensions, applicable standard and power supply requirements.
A: Yes. The UP-6195 is supplied with a 60 L work chamber as standard, and the internal dimensions / volume can be customized.
A: Yes. A Ø50 mm test port on the left side allows external power and signal cables to pass into the work chamber.
A: The independent over-temperature limit alarm automatically interrupts operation to protect the experiment and the samples.
A: No. Both ranges are configurable on request.



Need a quotation or a customized UP-6195 configuration? Contact us with your required temperature and humidity ranges, sample size, test standard and power supply, and our team will recommend the right chamber setup for your application.